model b-line v - southeastern...
TRANSCRIPT
Model b-line VZüllig Transmitter
Instruction ManualPN 51-B-LINEV/rev.B
February 2006
Emerson Process Management
Liquid Division2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2006
WARNING
ELECTRICAL SHOCK HAZARD
Making cable connections to and servicing
this instrument require access to shock haz-
ard level voltages which can cause death or
serious injury.
Be sure to disconnect all hazardous voltage
before opening the enclosure.
Relay contacts made to separate power
sources must be disconnected before servic-
ing.
Electrical installation must be in accordance
with the National Electrical Code
(ANSI/NFPA-70) and/or any other applicable
national or local codes.
Unused cable conduit entries must be
securely sealed by non-flammable closures
to provide enclosure integrity in compliance
with personal safety and environmental pro-
tection requirements.
The unused conduit openings need to be
sealed with NEMA 4X or IP65 conduit plugs
to maintain the ingress protection rating
(IP65).
For safety and proper performance this
instrument must be connected to a properly
grounded three-wire power source.
Proper relay use and configuration is the
responsibility of the user.
No external connection to the instrument of
more than 69VDC or 43V peak allowed with
the exception of power and relay terminals.
Any violation will impair the safety protection
provided
Do not operate this instrument without front
cover secured. Refer installation, operation
and servicing to qualified personnel.
ESSENTIAL INSTRUCTIONSREAD THIS PAGE BEFORE PROCEEDING!
Rosemount Analytical designs, manufactures, and tests itsproducts to meet many national and international stan-dards. Because these instruments are sophisticated tech-nical products, you must properly install, use, and maintainthem to ensure they continue to operate within their normalspecifications. The following instructions must be adheredto and integrated into your safety program when installing,using, and maintaining Rosemount Analytical products.Failure to follow the proper instructions may cause any oneof the following situations to occur: Loss of life; personalinjury; property damage; damage to this instrument; andwarranty invalidation.
• Read all instructions prior to installing, operating, and
servicing the product. If this Instruction Manual is not the
correct manual, telephone 1-949-757-8500 and the
requested manual will be provided. Save this Instruction
Manual for future reference.
• If you do not understand any of the instructions, contact
your Rosemount representative for clarification.
• Follow all warnings, cautions, and instructions marked
on and supplied with the product.
• Inform and educate your personnel in the proper instal-
lation, operation, and maintenance of the product.
• Install your equipment as specified in the Installation
Instructions of the appropriate Instruction Manual and
per applicable local and national codes. Connect all
products to the proper electrical and pressure sources.
• To ensure proper performance, use qualified personnel
to install, operate, update, program, and maintain the
product.
• When replacement parts are required, ensure that qual-
ified people use replacement parts specified by
Rosemount. Unauthorized parts and procedures can
affect the product’s performance and place the safe
operation of your process at risk. Look alike substitu-
tions may result in fire, electrical hazards, or improper
operation.
• Ensure that all equipment doors are closed and protec-
tive covers are in place, except when maintenance is
being performed by qualified persons, to prevent electri-
cal shock and personal injury.
TransmitterInstruction Manual
Transmitter b-line II
Installation, Connection, Start-up, Opera-tion, Calibration
1318e02 22/02/2005
Züllig AG • CH-9424 Rheineck • Phone +41(0)71 886 91 11 • Telefax +41(0)71 886 91 66
Instrumentation for waterand waste water control
2 Instruction Manual
Structure
Instruction manual (this document)Instructions for installation and ele-mentary operation information
Calibration instructions and menu treeSupplementary information concerningcalibration of the probe, as well asmeasurement specific menu tree.
User’s guideReference book with explanations anddescriptions of the transmitterfunctions.
Additional manuals- Housings- Auxiliaries- Options
Use of the technical documentation
Use
‹– Installiation‹– Start up‹– Operation‹– General menu
‹– Calibration of measurement chain‹– Measurement specific informations‹– Measurement specific menu structure
‹– term explanation‹– Functionswith search index!
‹– Design‹– Installiation‹– Functions
with search index –›
Subject to ChangeIn the interest of our customers, we reservethe right to make changes due to advance-ments in the state of technology.Illustrations, descriptions and contents ofdelivery are therefore not binding!
software versionThis document is valid for software versionSW 6101xx
Instruction Manual
Contents
3
Use of the technical documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Advice for the use of instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
For your safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Preparations for start up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Intended purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Control delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Electric connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
First start up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Start up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Short instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Additional information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Menu tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Menu diagnosis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Wiring diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4
Advice for the use of instructions
Transmitter b-line multi / b-line II / b-line
Helps for the settings
Complete, measurement specific menu overview
Instruction maual Calibration Instruc-tions
ApplicationNecessary documents:- Instruction manual Transmitter b-line- Calibration Instructions Transmitter b-line
When opening the menu pages of theinstruction manual and the calibrationinstructions, place the brochures one besidethe other in order to see the completemeasurement specific menu overview.- In the instruction manual you’ll find
general settings. Those are the same for allmeasurement types
- The calibration instructions contain thecorresponding menu structure for eachmeasurement, as well as calibration in-structions. Choose the double-sided menucorresponding to your measurement type.
Explanations of terms and functions aregiven in the user’s guide.
Note for probe connectionNecessary documents:- Instruction manul of transmitter- Instruction manual of probe
Proceed the same way as explained above.
General menu points Measurement specific settings
5 Important
For your safety
Please take notice of the following notes for your safety
Qualified personnelQualified personnel within the meaning ofthis instruction manual are persons familiarwith installing, starting-up, and operatingthis product, and who possess the qualifica-tion adequate to their work. Qualifiedpersons are:a) Educated and authorised to ground,
label, and switch on and off circuits,devices and systems according to thecurrent standards of safety technology.
b) Educated in maintenance and use ofadequate safety equipment according tothe current standards of safety technolo-gy.
c) Trained in first aid.d) For sites with explosion proofness:
Educated, trained and authorised towork at electrical switching circuits ofpotentially explosive sites .
RegulationsSpecific rules and laws in the country of theuser must be obeyed by the user.
Limits of liabilityThis warranty shall be null, void andinoperative in case of other than the inten-ded use, in case of reparation or alteration,as well as in case of inexpert use or incorrecthandling.
ApplicationThe unit may exclusively be used accordingto its intended purpose. It is equippedaccording to the customers indications andhas been tested before delivery. The personinstalling has to test the configuration andreport differences to the distributor. Instal-lation and connection has to be carried outby qualified personnel.
Attention!19“-inserts may only be used in completelyclosed protected type housings according tothe regulation. Never make alive when open!There are alive components on the printedcircuits.
Electrical InstallationTurn off current supply before installation ofthe unit. 230/115 VAC versions requireincreased safety measures. The units may onlybe installed and started up by qualifiedpersonell.
Overvoltage protection«Overvoltage protection» is the name forthe electrical protective arrangementsagainst atmospheric caused overvoltages atthe units.The filters in the unit are effective forovervoltages up to 2 kV (fine overvoltageprotection). For better protection please usethe Züllig overvoltage protection.
6 General
Preparations for start up
First start up
1. Make sure all connections and supplypower correspond to the unitsrequirements.
2. Turn on supply power.3. Start the unit up following the
instructions under «Start up».
Note: descriptions of functions!Detailed descriptions of the extensivefunctions can be found in the user’sguide.
Intended purpose
The Züllig transmitters are designed forevaluation of probe measuring values andfor triggering probe functions.The units may not be used in potentionallyexplosive areas. It is absolutely necessary tofollow the instructions in the instructionmanual!
Control delivery
The units are strictly tested before deliveryand are certain to leave the factory inperfect state. Control if the delivery iscomplete and check if there are damagescaused by transport.• Unpack unit• Do not throw away the original packing• Handle carefully• Do not touch any electronic components
Electric connection
Carry out the electric connections follo-wing the attached wiring diagram as perthe actual regulations. Take notice of thesafety information!Never manipulate at the open unit! Beforeany kind of manipulation at the unitsconnections cut off supply voltage!
Note concerning the probe connection!Make sure that probe and transmittermeasuring type match (this is particularlyimportant when different measuring typesare used at the same time)! Disregard candamage or distroy probe and transmitterparts.
7
Start up
Transmitter b-line/b-line II
Setting parameters
1. Units UNITDefault-settings can be changed ifnecessary.
2. Calibration CALCalibrate measuring chain as percalibration instructions.
3. Analogue outputs I-OUT (1 and 2)Is set to 4…20 mA by manufacturer.Change if necessary.
4. Integration time T-INTIf necessary set integration time(recommended).
5. Relay outputs REL (1 and 2)When using the relay it is necessary topreselect its type, and if it is used aslimit value relay activation, deactivati-on and dead time are to be set.
Short instructions Additional information
UnitsThere are different measurement specificunits available. It is possible to assigndifferent units to the measured values.Change the settings in case you do notwant to use the default settings.
CalibrationCalibration puts the probe signal in relati-on to the measured variables to be givenout. Calibration is measurement specificand is described in the separate calibrationinstructions.
Analogue outputsThe measured main value (b-line II also asecond value) can be given out as analoguesignal to external devices. There are twosignal ranges with 4…20 mA and 0…20mA available.
Integration timeSet the integration time for smoothing theprobe measuring value.
Relay outputsThe following relay type are available:For REL1:• Alarm relay• Limit value relay for main value• Limit value relay for supplementary valueFor REL2:• Counter impulse relay• Limit value relay for main value• Limit value relay for supplementary valueNotes:- It is possible to individually set activation
and deactivation point as well as dead timefor the limit value relays.
- The counter impulse relay is only availablefor flow measurement. It is possible to setthe quantity for the assigned counter unitper impulse for the counter impulse relay.
Use the menu tree (atthe back of thismanual) as help forfinding the programlevels for the setting.
8
I-OUT 1‹- TYPE Enter‹- MIN Enter‹- MAX Enter‹- LIV I1 Enter
Menu tree
General Settings
TEST‹- MODE Enter‹- AMODE Enter‹- SIMUL Enter‹- REL 1 T Enter‹- REL 2 T Enter‹- IOUT 1T Enter‹- IOUT 2T Enter
REL 1‹- TYPE Enter‹- ON2) Enter‹- OFF2) Enter‹- DTIME2) Enter‹- AL-DT3) Enter2) only Lim-MV/Lim-SV3) only with type Alarm
REL 2‹- TYPE Enter‹- 1 IMP1) Enter‹- ON2) Enter‹- OFF2) Enter‹- DTIME2) Enter1) only for C-Imp2) only Lim-MV/Lim-SV
T-INT‹- T-INT Enter
Terms and symbols
Unit
The parameter [unit] stands for an individual unit. Thisunit can be globally preselected for the correspondingmeasured variable in the «UNIT» menu .
4-20 + 0-20 Entervalue input (with ‹- and +) Entervalue input (with ‹- and +) EnterOFF + LOW + HIGH Enter
SYSTEM
CALsee measurement spe-cific calibration instruc-tions
COM‹- TYPE Enter‹- DEVNR Enter‹- BAUD Enter‹- PARITY Enter‹- WAIT Enter
UNITsee measurement spe-cific calibration instruc-tions
UNIT
SYSTEM
+
Enter
0
‹- CAL Enter‹- TEST Enter‹- I-OUT 1 Enter‹- I-OUT 2 Enter‹- REL 1 Enter‹- REL 2 Enter‹- T-INT Enter‹- COM Enter‹- UNITS Enter
CAL
Enter
Use (+) key to select ac-tion type:0 = read/control parame-
ter (Default)1 = set parameter
Leave main menu (dis-play modus) by pressingEnterkey
Activate Input menu bypressing Enter key
value input (with ‹- and +) Enter
C-Imp + Lim-MV + Lim-SVvalue input (with ‹- and +)1) Entervalue input (with ‹- and +)2) Entervalue input (with ‹- and +)2) Entervalue input (with ‹- and +)2) Enter1) only for C-Imp2) only Lim-MV/Lim-SV
Profib + Rackb + S-Bus Entervalue input (with ‹- and +) Enter300 + .... + 19200 EnterNo + Odd + Even EnterNo + 20ms + 100ms + 500ms Enter
1
2,5
normal + HOLD + SIMUL EnterHOLD + SIMUL + NoHold Entervalue input (with ‹- and +) EnterAuto + ON + OFF EnterAuto + ON + OFF Entervalue input (with ‹- and +) Entervalue input (with ‹- and +) Enter
I-OUT 2 1)
‹- TYPE Enter‹- MIN Enter‹- MAX Enter‹- LIV I2 Enter1) only b-line II
4-20 + 0-20 Entervalue input (with ‹- and +) Entervalue input (with ‹- and +) EnterOFF + LOW + HIGH Enter
Alarm + Lim-MV + Lim-SV Entervalue input (with ‹- and +)2) Entervalue input (with ‹- and +)2) Entervalue input (with ‹- and +)2) Entervalue input (with ‹- and +)3) Enter2) only Lim-HW/Lim-NW3) only with type Alarm
Menu tree symbols
‹- 1 line down in the menu or1 decimal to the left in the numerical valueinput
+ 1 position to the right in the value input or add1 to numerical value
Clear Return (present value remains)Enter Activation of menu points respectively confir-
mation of actual setting
9
UNIT
2,5
Menu Diagnosis
UNIT
Menu supp. value (notalways available)go on with arrow key
19,5
Display main value
Display supp. value
Menu probe diagnosis(not always available)go on with arrow key
Display probe diagnosis
Menu Transmitter Infor-mations (not alwaysavailable)go on with arrow key
Display transmitterinformations
Clear
Back with Clear
Note:The Values of display• Supp. Value• Probe diagnosis• Transmitter informationare different refer to measuring type
See also calibrating instructions
General Settings
Leave main menu (dis-play modus) by pressingarrow key
10
Wiring diagram
Transmitter b-line II
Wiring positions Important notes1) Protect relay outputs with RC-modules!
Do not exceed burden for contacts as pertechnical data!
2) Check supply voltage in accordance withthe device voltage
Connector box
Connector position in the connector box
Connectors
Clamp positions
Option with socket instead of screwed cable glandsfor the connection of the probe
Fuse
F 1 at 230 VAC 80 mATF 1 at 115 VAC 160 mATF 1 at 24 VDC 630 mAT
1)
1)
In2(+)In1(+)
A2B2
Ground24V(+)
K2K2K1K1
1098
Iout2(+)Iout (–)Iout1(+)
48 VAC248 VAC1
65 4 3 2 1
PE(+) L(–) N
PE
b-line II / b-line V
~~
~~
RC 1)
RC 1)
+ –A
+ –A
1
2
3
4
5
6
8
9
10
PE
L (+)
N (–)
PE
48V
AC
K1
K2
10 analogue output 29 analogue Ground8 analogue output 1
supply probe motor
6 probe input 25 probe input 14
interface RS48532 probe ground1 probe supply
K2relay contact 21)
K2K1
relay contact 11)
K1
L (+)N (–) Supply2)
PE
12
TransmitterdataDisplay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . LCD display
measuring value: 6 digits 7 segmentsmenu: 6 digits 16 segmentsand various special signs
Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Züllig probes of the 25-line with the excepti-on of PHKP-25 and RX-25;oxygen probe S-14;digital interface RS485 for sonic sensors;turbidity/dried solids COSMOS®-25
Analog Outputs . . . . . . . . . . . . . . . . . . . . . . 2 current outputs 0/4...20 mA,max. loading 600 ohm
Relay Outputs . . . . . . . . . . . . . . . . . . . . . . . . REL 1: potential free double limit switches asalarm or as limiting valueREL 2: potential free double limit switches aslimiting value or for measuring types EA/NA as counter
Contact Loading . . . . . . . . . . . . . . . . . . . . . . max. 2 A, 230 VACPower Supply Voltage . . . . . . . . . . . . . . . . . . 230 VAC +10/-15 %, 50/60 Hz (Standard)
115 VAC +10/-15 %, 50/60 Hz (Option)24 VDC +10/-15 % (Option)
Power Input . . . . . . . . . . . . . . . . . . . . . . . . . max. 14 VA + heating < 16 W (no heatingfor 24 VDC version)
Interfaces (Option) . . . . . . . . . . . . . . . . . . . . RS232: protocol Züllig Rackbus or SAIA SBus DatamodeProfibus: Slave DP V1 with PA profile,version 3
Transmission Distance . . . . . . . . . . . . . . . . . . max. 300 m (probe to transmitter)Protection Class . . . . . . . . . . . . . . . . . . . . . . 2Temperature Range . . . . . . . . . . . . . . . . . . . . -20...50 °CMoisture Range . . . . . . . . . . . . . . . . . . . . . . . max. 90 % r.M. not condensatingEnclosure . . . . . . . . . . . . . . . . . . . . . . . . . . . IP 66 (incl. heating)Connection . . . . . . . . . . . . . . . . . . . . . . . . . clamps 2.5 mm2
Technical Data
Transmitter b-line II
Power consumption 24 VDC versionPower for Transmitter (without heating) with connected Probe:with COSMOS®-25 (without wiper) . . . . . . . 4.6 Wwith other -25 Probes or presso/EXpresso . . 2.4 Wwith HT-68 or LT-68 . . . . . . . . . . . . . . . . . . 3.0 W
Profibus . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1.2 W (additional)
Instrumentation for waterand waste water control
TransmitterCalibration Instructions
Transmitter b-line II and b-lineOX, OH, DS/TM, PHBL/PHZ/PHDE, RXBL, LF, EA/EN/EV, NA/NM/NVMeasurement
Measurement Specific Information about:Calibration, Diagnosis, Menu Structure
1267e10 22/02/2005
Züllig AG • CH-9424 Rheineck • Phone +41(0)71 886 91 11 • Telefax +41(0)71 886 91 66
2
Subject to ChangeIn the interest of our customers, we reservethe right to make changes due to advance-ments in the state of technology.Illustrations, descriptions and contents ofdelivery are therefore not binding!
Calibration Instuctions Transmitter b-line II/b-line
Calibration Instructions Transmitter b-line II/b-line
Contents
3
Menu tree DO - Dissolved oxygen
Menu tree DS - Dried solids (total suspended solids)
Menu tree TR - Turbidity
Calibration OX/DO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Calibration for DS/TM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Calibration for DS/TM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
General settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Calibration journal and parameters list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Analysis measurement:
DO Oxygen probe S-12
DS Dried solid matter measurement with COSMOS® -25 probe
TM Turbidity measurement with COSMOS® -25 probe
Transmitter measuring type with corresponding probe!
MA
IN M
EN
UD
O:
Dis
so
lved
Oxyg
en
O2
Valu
e
SY
ST
EM
0D
ISL
AY
IN
FO
RM
AT
ION
Rea
d Pa
ram
eter
sU
se key
CA
LT
ES
TI-
OU
T 1
I-O
UT
2R
EL
1R
EL
2T
-IN
TC
OM
UN
ITS
TE
MP
Cal
ibra
tion
Mea
sure
men
t or A
larm
Set
tings
Anal
og o
utpu
t 1: D
OAn
alog
out
put 2
: Tem
pR
elay
1U
nits
Tem
pera
ture
VA
LU
EM
OD
ET
YP
ET
YP
ET
YP
ET
YP
ET
-IN
TT
YP
EO
2S
AT
DO
val
ueM
easu
rem
ent
Out
put r
ange
Out
put r
ange
Rel
ay T
ype
Rel
ay T
ype
Inte
grat
ion
Tim
eSe
rial C
omm
unic
atio
nD
isso
lved
Oxy
gen
O2
Satu
ratio
n
Ente
r X
X v
alu
e
Sele
ct N
orm
al,
Hold
or
Sim
ul
Sele
ct 0-2
0 o
r 4-2
0
Sele
ct 0-2
0 o
r 4-2
0
Sele
ct A
larm
, Lm
Tem
p, Lm
O2,
Lm
SA
T, Lm
SS
I
Sele
ct Lm
Tem
p, Lm
O2, Lm
SA
T, Lm
SS
IE
nte
r X
Seconds
Sele
ct R
ackb, S
-Bus,
Pro
fibus
Sele
ct m
g/L
or
ppm
SS
I
P-T
YP
EA
MO
DE
MIN
MIN
ON
(1)
ON
(1)
DE
VN
RS
AT
O2
Satu
ratio
n In
dex
Elec
trode
type
Alar
m
Min
imum
DO
Val
ueM
inim
umTe
mp
Valu
eO
nO
nD
evice
Num
ber
O2
satu
ratio
nF
e o
r Z
nS
ele
ct H
old
, S
imul o
r N
o H
old
Ente
r M
inim
um
valu
ve
Ente
r M
inim
um
valu
eE
nte
r O
n V
alu
eE
nte
r O
n V
alu
eE
nte
r V
alu
e (
00 -
89)
Sele
ct m
g/L
or
ppm
SL
OP
E
X.X
SA
LS
IMU
L 1
MA
XM
AX
OF
F (1
)O
FF
(1)
BA
UD
SS
I
Salin
itySi
mul
ate
DO
M
axim
um D
O V
alue
Max
imum
Tem
p Va
lue
Off
Off
Baud
Rat
eO
2 Sa
tura
tion
Inde
xE
RR
OR
Ente
r X
X v
alu
eE
nte
r X
X D
O v
alu
eE
nte
r M
axim
um
valu
eE
nte
r M
axim
um
valu
eE
nte
r O
ff V
alu
eE
nte
r O
ff V
alu
eS
ele
ct R
ate
Sele
ct %
or
bla
nk
Erro
r Num
ber
EL
EV
SIM
UL
2D
TIM
E (1
)D
TIM
E (1
)P
AR
ITY
(3)
TE
MP
S
ER
-Nr
Elev
atio
nSi
mul
ate
Tem
pD
elay
Tim
eD
elay
Tim
eD
ata
Parit
yTe
mpe
ratu
reAn
alyz
er S
eria
l Num
ber
Ente
r X
X F
tE
nte
r X
X tem
p v
alu
eE
nte
r X
seconds
Ente
r X
seconds
Sele
ct N
o, E
ven, O
dd
Sele
ct C
, K
, F
,or
R
SW
TE
MP
RE
L 1
TA
L-D
T (2)
WA
ITSo
ftwar
eV e
rsio
nTe
mpe
ratu
reR
elay
1Al
arm
Del
ayTi
me
Betw
een
Req
uest
sE
nte
r X
X T
em
pS
ele
ct A
uto
, O
n, O
ffE
nte
r X
seconds
Sele
ct m
s V
alu
eS
TA
RT
S
Num
ber o
f Sta
rtsR
EL
2 T
Rel
ay 2
HO
UR
S
Sele
ct A
uto
, O
n, O
ffH
ours
Ope
ratio
n
IOU
T 1
T
Anal
og o
utpu
t 1: D
O %
Ente
r valu
e
IOU
T 2
T
Anal
og o
utpu
t 1: T
emp
%E
nte
r valu
e
(1)
Only
availa
ble
with
TY
PE
: Lm
Tem
p, Lm
O2, Lm
SA
T, Lm
SS
I
(2)
Only
availa
ble
with
TY
PE
: A
larm
(3)
Only
availa
ble
for
Rackbus a
nd
S-b
us
Leg
en
d
1st lin
eB
OL
DA
naly
zer
Dis
pla
y
2nd li
ne
Italic
sD
efin
ition
3rd
line
Text
Optio
ns
Feb, 2006
QU
ICK
RE
FE
RE
NC
E G
UID
E
ME
NU
TR
EE
FO
R Z
ÜL
LIG
DIS
SO
LV
ED
OX
YG
EN
(D
O)
SY
ST
EM
Rea
d or
Cha
nge
Para
met
ers
SY
ST
EM
1
MA
IN M
EN
UD
S:
Dri
ed
So
lid
s (
To
tal S
usp
en
ded
So
lid
s)
DS
Valu
e
SY
ST
EM
0D
ISL
AY
IN
FO
RM
AT
ION
Rea
d Pa
ram
eter
sU
se key
CA
LT
ES
TI-O
UT
1I-
OU
T 2
RE
L 1
RE
L 2
T-I
NT
CO
MU
NIT
SD
S
Cal
ibra
tion
Mea
sure
men
t or A
larm
Set
tings
Anal
og o
utpu
t 1: D
SAn
alog
out
put 2
: DS
Rel
ay 1
Uni
tsD
ried
Solid
s
ME
MO
RY
MO
DE
TY
PE
TY
PE
TY
PE
TY
PE
T-I
NT
TY
PE
DS
CH
EC
K %
Rec
ord
Poin
tsM
easu
rem
ent
Out
put r
ange
Out
put r
ange
Rel
ay T
ype
Rel
ay T
ype
Inte
grat
ion
Tim
eSe
rial C
omm
unic
atio
nD
ried
Solid
s (T
SS)
Stan
dard
Dev
iatio
n
Sele
ct N
orm
al,
Hold
or
Sim
ul
Sele
ct 0-2
0 o
r 4-2
0
Sele
ct 0-2
0 o
r 4-2
0
Sele
ct A
larm
, Lm
DS
Lm
DS
(D
efa
ult)
Ente
r X
Seconds
Sele
ct R
ackb, S
-Bus,
Pro
fibus
Sele
ct ppm
, m
g/L
, g/L
, %
, (-
)
PS
ER
Nr
P1 (1
)A
MO
DE
MIN
MIN
ON
(2)
ON
(2)
DE
VN
RSe
nsC
alib
ratio
n Pt
1Al
arm
M
inim
um D
S Va
lue
Min
imum
DS
Valu
eO
nO
nD
evic
e N
umbe
rE
nte
r V
alu
eS
ele
ct H
old
, S
imul o
r N
o H
old
Ente
r M
inim
um
valu
ve
Ente
r M
inim
um
valu
eE
nte
r O
n V
alu
eE
nte
r O
n V
alu
eE
nte
r V
alu
e (
00 -
89)
P-S
W
Sens
or S
oftw
are
Num
ber
P2 (1
)S
IMU
L 1
MA
XM
AX
OF
F (2
)O
FF
(2)
BA
UD
Cal
ibra
tion
Pt 2
Sim
ulat
e D
SM
axim
um D
S Va
lue
Max
imum
DS
Valu
eO
ffO
ffBa
ud R
ate
PS
TA
RT
Ente
r V
alu
eE
nte
r X
X D
S v
alu
eE
nte
r M
axim
um
valu
eE
nte
r M
axim
um
valu
eE
nte
r O
ff V
alu
eE
nte
r O
ff V
alu
eS
ele
ct R
ate
Num
ber o
f sen
sor s
tarts
P3 (1
)R
EL
1 T
DT
IME
(2)
DT
IME
(2)
PA
RIT
Y (
3)
Cal
ibra
tion
Pt 3
Rel
ay 1
Del
ay T
ime
Del
ay T
ime
Dat
a Pa
rity
Erro
r Num
ber
Ente
r V
alu
eS
ele
ct O
ff, O
n, A
uto
Ente
r X
seconds
Ente
r X
seconds
Sele
ct N
o, E
ven, O
dd
SE
R-N
r
Dele
teR
EL
2 T
AL
-DT
(3)
WA
ITA
naly
z
Del
ete
Poin
tsR
elay
2Al
arm
Del
ayTi
me
Betw
een
Req
uest
s?
Sele
ct O
ff, O
n, A
uto
Ente
r X
seconds
Sele
ct m
s V
alu
eS
W
Anal
yzer
Sof
twar
e N
umbe
rW
IPE
RIO
UT
1T
Wip
er O
pera
tion
Anal
og o
utpu
t 1: D
S %
ST
AR
TS
Turn
On o
r O
ffE
nte
r valu
eN
umbe
r of S
tarts
T-W
IPIO
UT
2T
HO
UR
S
Wip
er In
terv
alAn
alog
out
put 1
: DS
%H
ours
Ope
ratio
nS
ele
ct 20 -
99 m
inute
sE
nte
r valu
e
(1)
These p
oin
ts a
re o
nly
availa
ble
if
the a
valu
e h
as b
een r
ecord
ed
(2)
Only
availa
ble
with T
YP
E: LM
-
DS
(3)
Only
availa
ble
with T
YP
E: A
larm
(4)
Only
availa
ble
for
Rackbus a
nd
S-b
us
Leg
en
d
1st lin
eB
OL
DA
naly
zer
Dis
pla
y
2nd lin
eIta
lics
Def
initio
n3rd
lin
eT
ext
Options
Feb, 2006
QU
ICK
RE
FE
RE
NC
E G
UID
E
ME
NU
TR
EE
FO
R Z
ÜL
LIG
TO
TA
L D
ISS
OL
VE
D S
OL
IDS
(D
RIE
D S
OL
IDS
) S
YS
TE
M
Rea
d or
Cha
nge
Para
met
ers
SY
ST
EM
1
MA
IN M
EN
UT
R:
Tu
rbid
ity
TR V
alue
SY
ST
EM
0D
ISL
AY
IN
FO
RM
AT
ION
Rea
d Pa
ram
eter
sU
se key
CA
LT
ES
TI-
OU
T 1
I-O
UT
2R
EL
1R
EL
2T
-IN
TC
OM
UN
ITS
TR
Cal
ibra
tion
Mea
sure
men
t or A
larm
Set
tings
Anal
og o
utpu
t 1: T
RAn
alog
out
put 2
: TR
Rel
ay 1
Uni
tsTu
rbid
ity
ME
MO
RY
MO
DE
TY
PE
TY
PE
TY
PE
TY
PE
T-I
NT
TY
PE
TR
CH
EC
K %
Rec
ord
Poin
tsM
easu
rem
ent
Out
put r
ange
Out
put r
ange
Rel
ay T
ype
Rel
ay T
ype
Inte
grat
ion
Tim
eSe
rial C
omm
unic
atio
nTu
rbid
itySt
anda
rd D
evia
tion
Sele
ct N
orm
al,
Hold
or
Sim
ul
Sele
ct 0-2
0 o
r 4-2
0
Sele
ct 0-2
0 o
r 4-2
0
Sele
ct A
larm
, Lm
TR
Lm
TR
(D
efa
ult)
Ente
r X
Seconds
Sele
ct R
ackb, S
-Bus,
Pro
fibus
Sele
ct F
NU
, N
TU
, E
BC
PS
ER
Nr
P1 (1
)A
MO
DE
MIN
MIN
ON
(2)
ON
(2)
DE
VN
R
Cal
ibra
tion
Pt 1
Alar
m
Min
imum
TR
Val
ueM
inim
um T
R V
alue
On
On
Dev
ice
Num
ber
Ente
r V
alu
eS
ele
ct H
old
, S
imul o
r N
o H
old
Ente
r M
inim
um
valu
ve
Ente
r M
inim
um
valu
eE
nte
r O
n V
alu
eE
nte
r O
n V
alu
eE
nte
r V
alu
e (
00 -
89)
P-S
W
Sens
or S
oftw
are
Num
ber
P2 (1
)S
IMU
L 1
MA
XM
AX
OF
F (2
)O
FF
(2)
BA
UD
Cal
ibra
tion
Pt 2
Sim
ulat
e TR
Max
imum
TR
Val
ueM
axim
um T
R V
alue
Off
Off
Baud
Rat
eP
ST
AR
T
Ente
r V
alu
eE
nte
r X
X T
R v
alu
eE
nte
r M
axim
um
valu
eE
nte
r M
axim
um
valu
eE
nte
r O
ff V
alu
eE
nte
r O
ff V
alu
eS
ele
ct R
ate
Num
ber o
f sen
sor s
tarts
P3 (1
)R
EL
1 T
D
TIM
E (2
)D
TIM
E (2
)P
AR
ITY
(3)
Cal
ibra
tion
Pt 3
Rel
ay 1
Del
ay T
ime
Del
ay T
ime
Dat
a P
arity
Erro
r Num
ber
Ente
r V
alu
eS
ele
ct O
ff, on, A
uto
Ente
r X
seconds
Ente
r X
seconds
Sele
ct N
o, E
ven, O
dd
SE
R-N
r
WIP
ER
RE
L 2
TA
L-D
T (3)
WA
IT
Wip
er O
pera
tion
Rel
ay 2
Alar
m D
elay
Tim
e Be
twee
n R
eque
sts
Turn
On o
r O
ffS
ele
ct O
ff, on, A
uto
Ente
r X
seconds
Sele
ct m
s V
alu
eS
W
Anal
yzer
Sof
twar
e N
umbe
rT
-WIP
IOU
T 1
T
Wip
er In
terv
alA
nalo
g ou
tput
1: T
R %
ST
AR
TS
Sele
ct 20 -
99 m
inute
sE
nte
r va
lue
Num
ber o
f Sta
rts
IOU
T 2
TH
OU
RS
Ana
log
outp
ut 1
: TR
%H
ours
Ope
ratio
nE
nte
r va
lue
(1)
Defa
ult
valu
es a
re 0
.006, 400.0
and 2
000.0
for
P1, P
2, P
3
(2)
Only
ava
ilable
with
TY
PE
: LM
-
TR
(3)
Only
ava
ilable
with
TY
PE
: A
larm
(4)
Only
ava
ilable
for
Rackbus a
nd
S-b
us
Leg
en
d
1st lin
eB
OL
DA
naly
zer
Dis
pla
y
2nd li
ne
Italic
sD
efin
ition
3rd
line
Text
Optio
ns
Feb, 2006
QU
ICK
RE
FE
RE
NC
E G
UID
E
Rea
d or
Cha
nge
Para
met
ers
SY
ST
EM
1
4
UNIT O2 Enter SAT Enter SSI Enter TEMP Enter
value input Enter Select electrode type Enter value input Enter value input Enter
CAL VALUE Enter P-TYPE Enter SAL Enter ELEV Enter
Calibration OX/DO
Measurement specific settings for OX/DO
Display informationAbbreviation Parameter Units MeasurementsO2 Main value mg/L, ppm DO OXTEMP Sec. value 1 °C, K, °F, °R DO OXSAT Sec. value 21) mg/L, ppm DO OXSSI Sec. value 22) %, without unit DO OX
Abbreviation Parameter Unit Measurements
SLOPE Slope (without unit) DO OXCH1 Channel1 mA — OXCH2 Channel2 mA — OX
Abbreviation Parameter Units Measurements
ERROR Transmitter Information error number DO OXSER-Nr Transmitter Information serial number DO OXSW Transmitter Information software number DO OXSTARTS Transmitter Information number of starts DO OX1) as per EN25814/ISO5814 2) as per DIN38408, part 23
Select unit Enter Select unit Enter Select unit Enter Select unit Enter
Units for OX/DO measurement
Calibration for OX/DO measurement
Menu tree
Procedure for maintenanceSee «General settings» at the end of thisbrochure!
Legend:VALUE Comparative valueP-TYPE Electrode Type Fe, ZnSAL Salt contentELEV Elevation
O2
OxygenSAT Oxygen saturationSSI Oxygen saturation indexTEMP Temperature
Output signals «frozen»!The output signals, I-OUT and LM (LC)are «frozen», as soon as [Cal ] is displayed.
Reactivation of measurement!After calibration press «Clear» until youget to the main menu!
5
˚C
Procedure
Prepare calibration1. Set transmitter into calibration mode as
follows:Leave screen Enter
[1] Set system mode to 1 Enter [CAL] Select Enter
Only for start-up: do the following basicsettings
[P-TYPE] Select Enter [FE] or [ZN] Select type Enter [SAL] Select Enter Set value of salt content Enter [ELEV] Select Enter Set value of altitude Enter
2. Take out the probe and clean followingthe user’s guide of the probe..
3. Produce oxygen saturated water (seeinformation in the right column).
4. Put the probe into the saturated waterand wait some minutes, until the probehas the same temperature as the liquid.
Calibration [VALUE] Select Enter
a) Confirm1) saturation value with Enter2)
orb) Give in comparative value
respectively saturation Set value Enter2)
Finish calibration1. Return the probe to the measurement
point in order to start again the measu-rement mode.
2. Re-activate the measurement at thetransmitter by:Return to [CAL] ClearReturn to main menu Clear
1) Gives the saturation value of actual «tem-perature», «salt content» and «altitude» as aproposal. This value can be changed if youdo not calibrate basing on saturation.
2) [OK] indicates a succsessful calibration[BAD SLOPE] indicates a failed calibrati-on.
Prepare
Finish
Calibration
Calibration OX/DO
CLEAR ENTER+
b-line
CLEAR ENTER+
b-line
CLEAR ENTER+
b-line
Calibration for OX/DO
Note to calibrationWhen ever possible calibration should bedone basing on the saturation index. Thesaturation can be calculated automatically(as per EN25814, ISO5814) basing on«temperature», «salt content» and «altitu-de». There must be real oxygen saturation!Alternatively it is possible to use anyreliable comparative value.
How to procuce oxygen saturatedwaterIn case the conductivity is smaller than200 µS/cm or not known, one teaspoon ofsalt is added to 10 liters drinking water.For application in sewage treatment plantsuse suspensionfree water from the sedi-mentation tank.a) Variant 1:
Bubble in air (e.g. compressed air orusing an aquarium pump) through anaquarium aeration stone
b) Variant 2 (makeshift):Quick spill from one bucket to theother 10- to 25-times.
8
Calibration for DS/TM
Measurement specific settings for DS/TM
Display informationAbbreviation Parameter Units MeasurementsDS Dried solid matter g/L, %, ppm, mg/L DSTR Turbidity FNU, NTU, EBC TM
Abbreviation Probe diagnosis Measurements
CHECK standard deviation TM DSPSERNo serial number of probe TM DSP--SW software number of probe TM DSPSTART number of probe starts TM DS
Abbreviation Transmitter information Measurements
ERROR error number TM DSSER-No serial number TM DSSW software number TM DSSTARTS number of starts TM DSHOURS hour meter TM DS
Calibration for DS/TM measurement
Menu tree
Procedure for maintenanceSee «General settings» at the end of thisbrochure!
Set unitsSee «General settings» at the end of thisbrochure
CAL MEMORY Enter P11) Enter P21) Enter P31) Enter WIPER Enter T-WIP Enter Dim-K2) Enter
1) This menu point isonly available if thecorrespondingcalibration point (P1,P2, P3) has beenrecorded with MEMO-RY or is alreadycalibrated
2) This menu point isonly available for TM
UNITSDS Enter select unit Enter
Units for DS measurement
UNITSTR Enter select unit Enter
Units for TM measurement
take of selection Enter value input Enter value input Enter value input Enter take of selection Enter value input Enter value input Enter
Legend:MEMORY record points (P1, P2, P3, Delete)Delete delete pointsP1, P2, P3 assign concentrationWIPER wiper (Off, On)T-WIP wiper intervalDim-K dim coefficient
DS dried solidsTM turbidity (measurement)TR turbidity (measurement value)
Output signals «frozen»!The output signals, I-OUT and LM (LC)are «frozen», as soon as [ Cal ] is displayed.
Reactivation of measurement!After calibration press «Clear» until you getto the main menu!
9
Calibration Procedure
Prepare calibration1. Set transmitter into calibration mode as
follows:Leave screen Enter
[1] Set system mode to 1 Enter [CAL] Select Enter
2. Take out the probe and clean followingthe user’s guide of the probe
3. Take out a probe for the determinationof the dried solid matter content in thelaboratory
Calibration (record actual value)1. Record actual value for point 1:
[MEMORY ] Select Enter [ P1] Select Enter
- Put probe into calibration liquid and mix continuously.Start calibration with Enter and keep on mixing, until [ WAIT] is no more displayed (ca. 15 s)!Note: Keep minimal distances of 7cmto walls and bottom when mixing.
2. If necessary record further calibrationpoints [ P2 ] and [P3 ] with their corre-sponding dried solid matter content.
Calibration (assign set point)Starting from [MEMORY ]:
[ ›P1 ]1) Confirm Enter Set value Enter
Assign further set points to points[ ›P2 ]1) and [ ›P3]1) the same way.
1) Only the recorded points are shown. Thesign › symbolises that the point has beenrecorded, but that there has not beenassigned any set point.
Finish calibration1. Return probe to the measurement point
in order to start again the measurementmode.
2. Return to main menu Clear
Preparation
Calibration DS/TM
Calibration for DS/TM
10cm
7cm7cm
COSMOS - 25
Calibration
Attention! Mix duringcalibration and keepminimal distances tothe wall.
Finish
Note to TM:The complete, normed curve of the probe isfactory set. If necessary the values of the setpoints can be modified. A modified curve isindicated with TR*(abbreviation parame-ter). With SYSTEM=75 you can return tofactory settings.
Change Calibration
Correcting a calibration pointLeave screen by pressing: Enter
Set [ 1 ] (System mode 1) EnterSelect [CAL] EnterSelect [MEMORY] Entere.g. select [P1] (point to correct) Enter
Give in new value EnterFinish correcting Clear
Clearing a calibration pointSame procedure as described above for«Correcting...» but setting the concentrati-on of the corresponding calibration point tothe value 0,00 . After confirming the zero-value the calibration point is cleared.
Rejecting of collected pointsLeave screen by pressing: Enter
Set [ 1 ] (system mode 1) EnterSelect [CAL] EnterSelect [MEMORY] EnterSelect [Delete] EnterNote:Only the new collected values (stored inthe cache memory) are deleted. Calibra-tion points with assigned set pointremain in memory!
Finish rejecting Clear
10
Calibration for DS/TM
Calibration for DS/TM
Further Settings
Turn on/off wiper modeCheck, if the probe has a wiper installed. Ifthere is a wiper installed, turn on therequested wiper mode.
1. Turn on wiper mode as follows:Leave screen Enter
[1] Set system mode to 1 Enter [CAL] Select calibration menu Enter [WIPER] Select Enter [On]/[Off ] Select Enter
2. Set wiper cycle: [T-WIP] Select Enter Give in cycle length Enter
3. Re-activate measurement at the trans-mitter by:Return to [CAL] ClearReturn to main menu Clear
Dim-Coefficient (only turbiditymeasurement)Leave screen by pressing: Enter
Set [ 1 ] (System mode 1) EnterSelect [CAL] EnterSelect [Dim-K] Enter
Give in value EnterNote:Only for COSMOS®-25 with softwareversion SW41043 or higher!The measured turbidity value is auto-matically reduced by the value of theDim-Coefficient. This value is automati-cally reset to 0 when the calibration ischanged or reset (standard = 0).When Dim-K 0, this is shown withTR+ .
Finish correcting Clear
Operation messages
Diagnosis and error messages see user’sguide b-line II, No. 1268e
11
Calibration DS/TM
Calibration for DS/TM
Glossary
MEMORYNormally it is sufficient when one point[P1] is collected. This point should be inthe upper third of the expected measuringrange. For high concentrations (e.g.digested sludge) two points are necessary.The program does automatically recognizethis and displays the message «CAL.INSUFFICIENT –› » together with theinformation + or – . Plus means, that asecond point with a higher concentrationhas to be calibrated, and minus meansthat a point with a lower concentrationhas to be collected.
P1…P3Calibration points are first collected (readactual value) under menu point [MEMO-RY] and then the set point [>P1…>P3]can be given in in a second step (result oflaboratory evaluation).Important:Calibration can only be done, after a setpoint has be assigned to each collectedcalibration point (marked with an arrow>).
DeleteAll collected (safed in cache) points aredeleted. Calibration is not deleted.
CHECK(display while reading) . Shows thestandard deviation of the measurements.Informs about quality of measuringsolution (homogeneity), not about qualityof measurement.
26
General settings
General settings
Set unitsLeave screen Enter
Set system mode to [1] Enter Select [UNITS] Enter Select e.g. [TEMP] Enter Select unit (e.g. [°C]) EnterReturn to [UNITS] ClearReturn to main menu Clear
Note:Dependent of menu settings, apply thebuttons several times or not even.
UNITS TEMP Enter select unit Enter
Procedure for maintenanceWhen maintenance word (e.g. cleaning ofprobe) is done, the transmitter wouldindicate a malfunction or extreme values. Inorder to prevent this, it is possible to givein the output value for the time of themaintenance work manually.
Procedure:Set menu [TEST] /[MODE] to [Hold] or[Simul]. Hold takes over and gives out theactual value, Simul gives the manuallydefined Simul value.Under [AMODE.] it is possible to define ifin case of active alarm the Hold or theSimul value should be given out. Afterfinishing the maintenance work it is neces-sary to reset [TEST]/[MODE] to [Nor-mal]. Afterwards set [Log] to «Yes» in orderto register the maintenance in the report.
27
Calibration journal and parameters list
General
Date Probe Calibration Value Visa
number solution
Description Value
TransmitterUser’s Guide
Transmitter b-line II, b-line V and b-line
Design and description of functions
1268e05 10/03/2005
Züllig AG • CH-9424 Rheineck • Phone +41(0)71 886 91 11 • Telefax +41(0)71 886 91 66
Instrumentation for waterand waste water control
2 User’s Guide Transmitter b-line II/b-line
Subject to ChangeIn the interest of our customers, we reservethe right to make changes due to advance-ments in the state of technology.Illustrations, descriptions and contents ofdelivery are therefore not binding!
3 User’s Guide Transmitter b-line II/b-line
Contents
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Calibration [CAL] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Basic settings [TEST] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Analogue output [I-OUT] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8Relays [REL1] and [REL2] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9System times [T-INT] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Communication [COM] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Units [UNITS] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11System [SYSTEM] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Diagnosis messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Standard functions b-line II/b-line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Expressions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
4 User’s Guide Transmitter b-line II/b-line
5 User’s Guide Transmitter b-line II/b-line
Index
Symbole[I-OUT] 8[TYPE] 10
AAbbreviations 16Alarm 7, 14Analogue output 8Arrow to the left 6
BBaudrate 10
CCalibration 7Clear 6Communication 10
Ddata transmission 10Device number 10Display 16
EEnter 6Error messages 14Expressions 16
FFront elements 6Front panel 6
GGlossary 16
HHOLD 7
IInterface 10Interface configuration 10
Interface Type 10
LLC 16limit contact 16LIV I1 8LIV I2 8LiveZero 8
Mmaintenance 7Measurement 16Measuring type 15Menu 16Message 14metering version 15
PParameter 16Parity 10Plus 6Probe signal 16
RResponse time 11
SSIMUL 7Structure 6
TTransmission 11
UUnit 16
VValue 16Vertical navigation 6
6 General
Design
StructureThe b-line Transmitter is a complete equip-ment including field monitoring enclosureand integrated power supply.
Front panelThe tough membrane with integrated keysis very resistant. It is insensitive and veryeasy to handle. For cleaning a cloth moi-stened with ordinary soapy water can beused.
Front elements(1) LCD(2) operation keys
Key functionsEnter
The enter key has various functions:a) Switch from measuring value display to
Login Menu (system).b) - Select sub-menu.
- Select parameter for setting.- Confirm parameter.
Clear
The clear key is used for leaving a menu orthe settings level without storing the set-tings.Arrow to the left
- Vertical navigation in the menu tree. Eachtype causes a jump down in the menustructure.
- This arrow key is used when giving innumbers for changing the cursor position.
- Jump in the main menu from main valueto supp. value to to probe diagnosis totransmitter informations.
Plus
- This key is used when giving in numbersfor increasing the digit by one (9 is follo-wed by 0).
- When giving in switch-points the nextswitch position is activated (Horizontalnavigation in the menu tree. Each typecauses a jump to the right in the menustructure).
2
1
7 Main Functions
Calibration [CAL]
By calibration a corresponding measuringvalue is generated out of an electric signal,or a parameter is calculated basing onanother one. (e.g. pH-value out of potenti-al and temperature).Depending of the measuring chainfunction there are different procedures forcalibration. For detailed instructions seemeasurement specific calibration instruc-tions.
Main Functions
Basic settings [TEST]
Transmitter settings for its reactions duringmaintenance or alarm.
Maintenance• Metering [MODE]• Alarm-Mode [AMODE]• Simul.1 [SIMUL1]• Simul.2 [SIMUL2]
MeteringNORMALHOLDSIMULTurn on continuous measurement with«ON», or hold the last measuring valuewith «HOLD» or give out the value whichhas been set for «SIMUL» as measuringvalue. Status of Hold-Function is indicatedon display.
Alarm-ModeHOLD SIMUL NoHoldPossible settings for the transmitters reac-tion to a malfunction. See also «errormessages».HOLDTransmitter goes into HOLD-mode andgives out the last valid measuring valueuntil the malfunction is cleared.SIMULTransmitter goes into simulation-mode andgives out the selected measuring value assimulated value until there comes in again aproper measuring signal.NoHoldFaulty probe signal go on in use.
Simul.Enter the values according required datas.
8
Analogue output [I-OUT]
I-OUT at b-lineI-OUT 1 and I-OUT 2 at b-line II
I-OUT1 I-OUT2
TYPE TYPE MIN MIN MAX MAX LIV I1 LIV I2
TYPE4…20 mA ‹- 0…20 mA
MINValue which should correspond to theminimal output signal.
20 mA
0/4 mA
20 mA
MinMin MaxMax
Set MIN value and MAX value for optimalscaling of the analogue output.
MAXValue which should correspond to themaximal output signal.
LIV I1 resp. LIV I2Alarm output by analogue output (LiveZe-ro configuration).
Menu 4…20 mA 0…20 mA Description
OFF – – LiveZero off1)
LOW 3.5 mA 0 mA Current at alarmHIGH 21.5 mA 21.5 mA Current at alarm1)alarm: current output corresponds to Hold or Simul
LiveZero works irrespective of AMODEand irrespective of the configuration ofREL1 (as an alarm relais or as a limitcontact). The alarm delay time AL-DT alsoapplies to LiveZero.
Main Functions
Main Functions
9
Relays [REL1] and [REL2]
Relay 1 is assigned by the manufacturer tothe ALARM function.
REL1 TYPE ON OFF DTIME AL-DT
TYPE - ALARMThe relay indicates error messages.
TYPE - Lm-MVLimit contact for main measuring value.The main measuring value is defined in themeasurement specific calibration instruc-tions.
TYPE - Lm-SVLimit contact for supplementary measuringvalue. The supplementary measuring values(if available) are defined in the measure-ment specific calibration instructions.
REL2
TYPE 1 IMP (counter only) ON OFF DTIME
TYPE - Lm-MVLimit contact for main measuring value.The main measuring value is defined in themeasurement specific calibration instruc-tions.
TYPE - Lm-SVLimit contact for supplementary measuringvalue. The supplementary measuring values(if available) are defined in the measure-ment specific calibration instructions.
Main Functions
Main Functions
TYPE - C-IMPAt flow measuring relay 2 may be set ascounter impulse relay.
The following input field is only available incombination with an activated alarm relay!
AL-DTIt is possible to set an alarm delay for relaycontact 1. The delay does only concern therelay contact. Other functions as for examp-le the activation of the alarm mode (Hold,Simul) and error messages are not delayed.Return without delay. Standard setting 5seconds.
The following input fields are only available inconnection with an activated limit contact!
ONInput field for the connecting point of therelay.
OFFInput field for the disconnecting point ofthe relay.
DTIMEInput field for the delay time of the relay.
The following input fields are only available inconnection with an activated counter impulserelay!
1 IMPFlow per relay impulse.
Delay time for theprevention of limitcontact fluttering.Only when the limitvalue is exceededlonger than the presetdelay time the corre-sponding limit contactrelay switches.Example: Switchingon/off of externaldevices as samplingdevice, flushinginstallations, etc.
10
System times [T-INT]
Setting of the integration time.Interval, during which probe data are inte-grated for smoothing
Smoothing of probe data
Main Functions
Communication [COM]
Settings for the serial interface, which isused for communication with peripheraldevices. This is an optional additionalprint, which is configured depending onorder.
Available options:- RS232 or RS485 with protocol Züllig
Rackbus and SAIA S-Bus Datamode.- Profibus with Profibus protocol
Interface configuration• TYPE• DEVNR• BAUD• PARITY• WAIT
Interface Type [TYPE]The options for TYPE are as follows:- Rackb Züllig Rackbus- S-Bus SAIA S-Bus- Profibus Profibus is automatically selec-ted when Profibusprint is plugged in. Seeadditional instructions «PROFIBUS»!
If TYPE is set up differently, the subsequentcommunication parameters are automaticallyset on default values.Therefore, always select the type first andonly then the other parameters.
Device number [DEVNR]Input field for the device number of thetransmitter within a net. Permissible devicenumbers are between 00 and 89.
Baudrate [BAUD]Selection of requested data transmission rate.
Parity[PARITY]Only relevant for Rackbus and S-Bus proto-col!Data transmission parity:no, even or odd.
Main Functions
100%
90%
T-INT
t
t
Input signal
Output signal integrated
11
Response time [WAIT]Selection window for the time between adata request from a higher system and theresponce of the transmitter.Transmission with 8 Bit long data blocks andstop bit.For detailed informations see additionalmanual «Communication».
Main Functions
Main Functions
Units [UNITS]
Set a unit for a parameter. This unit is fromthen on used on all screens with the corre-sponding parameter.
There is a list of the measurement specificparameters with the corresponding units inthe menu tree chapter in the calibrationinstructions.
12 Main Functions
Main Functions
Counter
Note:With SYSTEM=900 you can set the COUN-TER at individual values.
13
System [SYSTEM]
The «SYSTEM» screens are accessible fromhere with a code. They are used for internaltesting and alignment of transmitter hard-ware.
These testing and alignment functions arereserved for instructed Züllig personnel andaccredited service centers.Warning!Improper settings at the internal parame-ters can cause malfunctions or failure ofthe transmitter. Züllig AG refuses any kindof responsibility for such maladjustedunits.
Access to menu:[SYSTEM]0: see menu parameters1: see or change menu parameters
Main Functions
Main Functions
Diagnosis messages
Informations about technical status oftransmitter as well as probe data.
Probe informationsSee measurement specific calibrationinstructions!
Transmitter informationsDisplay of informations regarding generaltransmitter functions.
ERROR
Shows error number of error message
SER-Nr
Shows serial number of hardware
SW
Shows version number of software
STARTS
Number of hardware starts
14
Error Messages
Error Messages
ExplanationError messages are caused by hardwareerrors and release the alarm relay. Operationscause messages. For example wrong inputsor calculation results outside the admissible
Text (light writing on display) Number Type Comment MeasureBAD SLOPE 2001 Message after bad calibration (untypical slope has been
calculated for the electrode)- new calibration (see also calibration instructions)- perhaps change electrode
BAD VENTURI PARAMETER 2005 Message Norm venturi dimensions outside range(only for RT)
- set right venturi parameter
CAL. INSUFFICIENT ->– 990012 Alarm further calibration point necessary - further calibration point with lower concentration necessaryCAL. INSUFFICIENT ->+ 990033 Alarm further calibration point necessary - further calibration point with higher concentration necessaryCLEANING PROBE 990003 Message probe dirty - clean probe
- perhaps caused by non-homogeneity of measuring liquidCOMMUNICATION PROBE 100001 Alarm no communication with digital probe - plug in probe
- check connection to probe- correct probe number
EEPROM 1 ERROR 300004 Alarm error in EEprom basic print (perhaps data loss) - turn instrument off (power off) and on again (power on)- if on repeated occasions have transmitter repaired
EEPROM 2 ERROR 300005 Alarm error in AD-converter basic print - turn instrument off (power off) and on again (power on)- if on repeated occasions have transmitter repaired
EEPROM ERROR 300001 Alarm error in EEprom CPU print (perhaps data loss) - turn instrument off (power off) and on again (power on)- if on repeated occasions have transmitter repaired
ERROR 1 PROBE 320210 Alarm no high-voltage - have echo-probe repairedERROR 2 PROBE 320220 Alarm no reference - clear space between echo head and reference bar
- clean reference bar- if cleaning does not help, have echo probe repaired
ERROR 4 PROBE 320240 Alarm no measuring echo - Echo probe must be installed exactly right-angled to the measuring surface.- check maximal measuring distance (echo to measuring surface) with data sheet of the probe.- if this does not help, have echo probe repaired
ERROR CAL.FUNCTION 990006 Alarm no or wrong curve available - new calibration necessary (there is no factory-set suspended solids curve in the probe)
ERROR PROBE 990088 Alarm EEprom in probe defective - turn instrument off (power off) and on again (power on)- if on repeated occasions have COSMOS-25 probe repaired
EXP. TOO LARGE 360300 Alarm Exponent too large - check formula inputMANTISSA ZERO 360400 Alarm Mantissa = 0 - check formula inputMANUFACTURER STANDARD 990000 Message over 4000 FNU(DIN-Norm only up to 4000FNU)NO SIGNAL CHANNEL 1 100010 Alarm no current signal from probe (main value) - plug in probe
- check connection to probe- DO: PT1000 has to be connected
NO SIGNAL CHANNEL 2 100020 Alarm no current signal from probe (temperature) - plug in probe- check connection to probe
OFFSET ERROR 990005 Alarm calibration gives no zero point - new calibration necessaryOK 2002 Message - after good calibration
- as confirmation for system code, not displayed on the menu
POINTS SAVED 2004 Message a concentration has been assigned to all collectede points
RANGE OVERFLOW 990001 Alarm - probe is measuring in inappropriated liquid -DS: perhaps new calibration necessary
WAIT 2003 Message record point (re-calibration) - wait (probe in calibration liquid)WAIT 90099 Alarm record point (first calibration, on the background there
is an error message as e.g. ERROR CAL. FUNCTION)- wait (probe in calibration liquid)
WRONG FORMULA 360200 Alarm wrong formula segment - check formula inputWRONG PZ 360100 Alarm check digit does not fit to formula segment - check formula input
Attention: not every measurement supports all error messages.
measuring range cause messages. Thedisplay disappears after some seconds. Errormessages and messages are shown ondisplay by scrolling.
AlarmTransmitter remains in HOLDrespectively SIMUL status, aslong as alarm reason persists.
-> alarm contact openMessageIs displayed during some seconds-> alarm contact closed
Error Messages/Messages
15
Functions
b-line II/b-line
Standard functions b-line II/b-line
Measuring typeFunction —›
Measuring chain (probe) ‹— M
eter
ing
vers
ion
Tw
o an
alog
ue o
utpu
ts3)
Tw
o re
lay
outp
uts
Uni
ts p
rese
lect
ion
Inte
grat
ion
tim
e
Line
ariz
atio
n
Cou
nter
Impu
lse
outp
uts
Analytical measurementOxygen, saturation, temperature metering (S-14) OXOxygen, saturation, temperature metering1) OHOxygen, saturation, temperature metering (S-12V) 2) DOpH metering (PHBL-25) PHBLpH and temperature metering (PHZ-/PHZB-/PHZB2-25) PHZRedox potential metering (RXBL-25) RXBLConductivity, temperature metering (LF-25) LF
Turbidity metering (COSMOS®-25) TMSuspended solids metering (COSMOS®-25) DSTemperature metering (external signal) TE
Flow, level and volume measurementLevel metering with echo sounder (HAT-68, LT-68) ENFlow, level metering with echo sounder (HT-68, LT-68) EAVolume, level metering with echo sounder (HT-68, LT-68) EV3)
Level metering with pressure sensor (presso) NMFlow, level metering with pressure sensor (presso) NAVolume, level metering with pressure sensor (presso) NV3)
1) available only in Switzerland, 2) special version, 3) only b-line II
16
Glossary
Expressions
MeasurementIndicates the transmitter function. Examp-le: OX for oxygen, LF for conductivity, RXfor redox potential etc.
ParameterPhysical value to be displayed or evaluated.Examples: Oxygen content, level etc.
UnitUnit for the corresponding parameterExamples: l/h, mm etc.
ValueActual value of a parameter with its corre-sponding unit
DisplayDisplay field with all informations on thedisplay, as texts, values, etc.
MenuThe menu is a list of possible selections foractivation of the requested functions.
Probe signalThe «basic measuring value» of the probe atthe transmitter input. This signal is conver-ted to a defined parameter basing on thecalibration.
TransmitterAmplifier
Glossary
Abbreviations
MV main value (on display)SV supplementary value (on display)
Displ. displayMIN minimal valueMAX maximal valueLC limit contactSIMUL measuring value simulationLCD liquid cristal display
WARRANTY
Goods and part(s) (excluding consumables) manufactured by Seller are warranted to be free from defects in workman-
ship and material under normal use and service for a period of twelve (12) months from the date of shipment by Seller.
Consumables, pH electrodes, membranes, liquid junctions, electrolyte, O-rings, etc. are warranted to be free from defects
in workmanship and material under normal use and service for a period of ninety (90) days from date of shipment by Seller.
Goods, part(s) and consumables proven by Seller to be defective in workmanship and / or material shall be replaced or
repaired, free of charge, F.O.B. Seller's factory provided that the goods, parts(s), or consumables are returned to Seller's
designated factory, transportation charges prepaid, within the twelve (12) month period of warranty in the case of goods
and part(s), and in the case of consumables, within the ninety (90) day period of warranty. This warranty shall be in effect
for replacement or repaired goods, part(s) and consumables for the remaining portion of the period of the twelve (12)
month warranty in the case of goods and part(s) and the remaining portion of the ninety (90) day warranty in the case of
consumables. A defect in goods, part(s) and consumables of the commercial unit shall not operate to condemn such com-
mercial unit when such goods, parts(s) or consumables are capable of being renewed, repaired or replaced.
The Seller shall not be liable to the Buyer, or to any other person, for the loss or damage, directly or indirectly, arising
from the use of the equipment or goods, from breach of any warranty or from any other cause. All other warranties,
expressed or implied are hereby excluded.
IN CONSIDERATION OF THE STATED PURCHASE PRICE OF THE GOODS, SELLER GRANTS ONLY THE ABOVE
STATED EXPRESS WARRANTY. NO OTHER WARRANTIES ARE GRANTED INCLUDING, BUT NOT LIMITED TO,
EXPRESS AND IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.
RETURN OF MATERIAL
Material returned for repair, whether in or out of warranty, should be shipped prepaid to:
Rosemount Analytical Inc.
Uniloc Division
2400 Barranca Parkway
Irvine, CA 92606
The shipping container should be marked:
Return for Repair
Model _______________________________
The returned material should be accompanied by a letter of transmittal which should include the following information
(make a copy of the "Return of Materials Request" found on the last page of the Manual and provide the following there-
on):
1. Location type of service, and length of time of service of the device.
2. Description of the faulty operation of the device and the circumstances of the failure.
3. Name and telephone number of the person to contact if there are questions about the returned material.
4. Statement as to whether warranty or non-warranty service is requested.
5. Complete shipping instructions for return of the material.
Adherence to these procedures will expedite handling of the returned material and will prevent unnecessary additional
charges for inspection and testing to determine the problem with the device.
If the material is returned for out-of-warranty repairs, a purchase order for repairs should be enclosed.
Credit Cards for U.S. Purchases Only.
The right people,the right answers,right now. ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
http://www.raihome.com
Specifications subject to change without notice.
Emerson Process Management
Liquid Division2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihome.com
© Rosemount Analytical Inc. 2006