Quantification TEMP/TDM Torsten Kuwert Clinique de Médecine Nucléaire

Upload: others

Post on 24-May-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Quantification TEMP/TDM

Torsten Kuwert

Clinique de Médecine Nucléaire

Page 2: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Quantification TEMP/TDM

Introduction

Quantification

Correction des artefacts

Calibration

Validation

Potentiel clinique

Résumé

Page 3: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

La lutte des technologies

TEMP TEP

Page 4: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

TEMP a des difficultés à mesurer

la concentration de la

radioactivité in vivo en unités

absolues, par exemple en kBq/ml

Progrès technologiques en TEMP?

Page 5: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

La chaîne de la formation des images TEMP

Ritt P et al. Eur J Nucl Med Mol Imag 2011

Les facteurs

perturbant

la mesure de la

C*

Page 6: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Quantification TEMP/TDM

Introduction

Quantification

Correction des artefacts

Calibration

Validation

Potentiel clinique

Résumé

Page 7: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Roadmap to Absolute Quantification with

SPECT/CT

True Activity

Distribution

Measured Activity

Distribution

SPECT Measured Count

Distribution

Calibration

Motion

correction

Attenuation

correction

Scatter

correction

Resolution

recovery

Reconstruction

Finite Resolution

Scatter

Attenuation

Dead Time

Motion

Page 8: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Le RC dépend du rapport entre le volume de la

structure d´intérêt et la résolution spatiale

2

Zeintl et al. JNM 2010

Page 9: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

La correction pour les artéfacts a été facilitée par

le développement de la reconstruction itérative

Corrections

Page 10: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

L´erreur de la quantification dépend du nombre

des itérations

Zeintl et al. JNM 2010

Page 11: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Recovery in Phantom

50%

55%

60%

65%

70%

75%

80%

85%

90%

0 32 64 96 128 160 192 224 256

# of Updates

Recocery

[%

]

Lu-177 113 keV 16ml

Lu-177 208 keV 16ml

Tc-99m 16ml

Lu-177 113 keV 8ml

Lu-177 208 keV 8ml

Tc-99m 8ml

Need to iterate more due to the lower

spatial resolution, compared to Tc-99m.

Le nombre nécessaire d´itérations dépend aussi

de l´isotope choisi

Page 12: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Roadmap to Absolute Quantification with

SPECT/CT

True Activity

Distribution

Measured Activity

Distribution

SPECT Measured Count

Distribution

Calibration

Motion

correction

Attenuation

correction

Scatter

correction

Resolution

recovery

Reconstruction

Finite Resolution

Scatter

Attenuation

Dead Time

Motion

Page 13: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

30 % de diffusé dans 140 keV± 5%

du spectre du Tc-99m

Correction de diffusé (soustraction)

Page 14: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Spectre en énergie du Lu-177

Lu-177 Spectra

1,E-05

1,E-04

1,E-03

1,E-02

1,E-01

1,E+00

0 50 100 150 200 250 300

Energy [keV]

Counts

(N

orm

alized)

HPGe

Page 15: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Lu-177 Spectra

1,E-05

1,E-04

1,E-03

1,E-02

1,E-01

1,E+00

0 50 100 150 200 250 300

Energy [keV]

Counts

(N

orm

alized)

HPGe NaI (T2 PointSource) NaI T2 (CylinderSource)

Spectre en énergie du Lu-177

Page 16: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Fenêtres en énergie

Page 17: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Roadmap to Absolute Quantification with

SPECT/CT

True Activity

Distribution

Measured Activity

Distribution

SPECT Measured Count

Distribution

Calibration

Motion

correction

Attenuation

correction

Scatter

correction

Resolution

recovery

Reconstruction

Finite Resolution

Scatter

Attenuation

Dead Time

Motion

Page 18: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Atténuation photonique

Fantôme cylindrique (In-111)

Page 19: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Coefficients d´atténuation linéaires vs.

unités d´Hounsfield

Conversion of HU to Linear Attenuation Coefficients

0,000

0,050

0,100

0,150

0,200

0,250

0,300

0,350

-100

0-9

00-8

00-7

00-6

00-5

00-4

00-3

00-2

00-1

00 010

020

030

040

050

060

070

080

090

0

1000

1100

1200

1300

1400

1500

Hounsfield Units

µ [

1/c

m]

µ(140 keV)

µ(364 keV)

µ(511 keV)

Page 20: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Image d´émission TDM „µ-map“

Correction de l´atténuation photonique par l´utilisation de la TDM

Page 21: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Homogénéisation de la distribution du traceur par la correction d´atténuation

Correction de l´atténuation photonique par l´utilisation de la TDM

Page 22: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Quantification TEMP/TDM

Introduction

Quantification

Correction des artefacts

Calibration

Validation

Potentiel clinique

Résumé

Page 23: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Roadmap to Absolute Quantification with

SPECT/CT

True Activity

Distribution

Measured Activity

Distribution

SPECT Measured Count

Distribution

Calibration

Motion

correction

Attenuation

correction

Scatter

correction

Resolution

recovery

Reconstruction

Finite Resolution

Scatter

Attenuation

Dead Time

Motion

Page 24: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Calibration de la caméra TEMP/TDM par rapport à un activimètre

Fantôme cylindrique contenant 200-300

MBq Tc-99m

Image courtesy of Zeintl J. 12,531 (cpm/ml)/(kBq/ml) on Symbia® T6

Page 25: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Tc-99m LEHR, Symbia T2: 5.62 cpm/kBq Lu-177 MELP, Symbia T2: 0.61, 0.65 cpm/kBq (f(keV))

Facteurs de calibration

Page 26: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Quantification TEMP/TDM

Introduction

Quantification

Correction des artefacts

Calibration

Validation

Potentiel clinique

Résumé

Page 27: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Quantitative Capabilities of four State-of-the Art SPECT-CT Cameras (3 Calibrations, 3 Phantoms, 5 ROIs)

20 %

- 25 %

Seret A et al. EJNMMI Res 2012

Error < 10% in large volumes

Page 28: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Le “gold standard”: La concentration de radioactivité dans la vessie et dans l´urine

16 patients

Page 29: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Zeintl et al. J Nucl Med 2010

Erreur moyenne: 6,8 %

(SE 8.0%; 95% CI: -14.6 – 16.6)

Quantification de C* à l´aide de la TEMP in vivo (Tc-99m)

Page 30: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Quantitative Lu-177-SPECT/CT

Validation in 8 Patients

0,00%

20,00%

40,00%

60,00%

80,00%

100,00%

120,00%

140,00%

1 2 3 4 5 6 7 8

Quantitative Accuracy in Bladder

Hand Segmentation Ellipsoid

In elliptical VOIs, mean error 8.03 %

Sanders J et al. (in prep.)

Page 31: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Premier outil pour la quantification TEMP

disponible commercialement

Carte de kBq/ml ou SUV

pixel par pixel

résultant directement

de la reconstruction

(9/2013,

XSPECT quant, Siemens)

SUV avg 6.2

Page 32: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Quantification TEMP/TDM

Introduction

Quantification

Correction des artefacts

Calibration

Validation

Potentiel clinique

Résumé

Page 33: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Quelle est la C* dans les corps vertèbraux

de patients soumis à la scintigraphie osseuse?

50 patients féminins

TEMP 4 ± 0,5 h après 8,26 ± 1,41 MBq DPD/ kg i.v.

Page 34: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

48,15 ± 13,66 kBq/ml

(95% CI 45.81 – 50.50; n = 50) (Cachovan et al. EJNMMI Res 2013)

C* dans les corps vertèbraux

de patients soumis à la scintigraphie osseuse

Page 35: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

5,91 ± 1,54 SUV

(95% CI 5,71 – 6,24) (Cachovan et al. EJNMMI Res 2013)

SUV dans les corps vertèbraux

de patients soumis à la scintigraphie osseuse

Page 36: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Potentiel clinique de la TEMP quantitative

Perturbations diffuses

de l´accumulation

d´un radiotraceur

Evaluation de la réponse

des tumeurs à la

thérapie

Page 37: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

SUV vs. Hounsfield Units dans un groupe de

50 femmes soumises à la DPD-TEMP/TDM

r = 0,678, p < 0,0001 (95% CI 0,575 – 0,761) Cachovan et al.

2013

Page 38: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

TEMP dynamique?

TEMP quantitative pourrait faciliter

l´analyse compartimentale

Blake GM et al. SNM 2001

ECF Plasma Bone

mineral

Krenal

K1

KBone K2

K4

Page 39: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Patient âgé de 75 ans atteint d´un cancer

de la glande prostatique (T2cpN0; Gleason 4 + 3 = 7)

PSA 01/13:

100 ng/ml

PSA 10/13:

395,6 ng/ml

Métastases

actives?

Page 40: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

SUV: 9 SUV: 9.5

Patient âgé de 75 ans atteint d´un cancer

de la glande prostatique (T2cpN0; Gleason 4 + 3 = 7)

Page 41: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Limites préliminaires de normalité

0.00

2.00

4.00

6.00

8.00

10.00

12.00

14.00

16.00

18.00

20.00

BG1

(L1)

BG2

(L2)

BG3

(L5)

BG4

(Pelvis)

BG5

(Pelvis)

BG6

(Pelvis)

T1 (L

3)

T2 (L

4)

T3 (P

elvis M

et.)

T4 (P

elvis M

et.)

SU

V

SUV Mean

SUV max

Lower 95% Conf.Int.

(Approx.)

Upper 95% Conf.Int.

(Approx.)

Patient âgé de 75 ans atteint d´un cancer

de la glande prostatique (T2cpN0; Gleason 4 + 3 = 7)

Page 42: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Tc-99m-HMPAO-SPECT du rCBF

En moyenne, % dose injectée de

2.18 ± 0.45 dans le cerveau entier (n = 32) (Welz et al. En préparation)

Page 43: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Global HMPAO Uptake corrèle avec la

performance cognitive

Cerebral %ID vs. MMSE: r = -0.34, p < 0.012

Page 44: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Potentiel clinique de la TEMP quantitative

Perturbations diffuses

de l´accumulation

d´un radiotraceur

Evaluation de la réponse

des tumeurs à la

thérapie

Page 45: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Female patient with abnormal uptake in skull

Patiente âgée de 58 ans atteinte

d´un cancer du sein (pT1c pN3a cM0; OP 03/2010)

Scintigraphie osseuse 03/2012

Foyer d´hyperfixation dans la calotte (SUV 4,8),

TDM normale

Recommendation d´un examen de suivi

Page 46: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Female patient with abnormal uptake in skull

Scintigraphie osseuse 6 mois plus tard

Augmentation du SUV de 4.8 à 6

Thérapie avec les bisphosphonates

Patiente âgée de 58 ans atteinte

d´un cancer du sein (pT1c pN3a cM0; OP 03/2010)

Page 47: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Female patient with abnormal uptake in skull

Scintigraphie osseuse encore 5 mois plus tard

Réduction du SUV de 6 à 5,1

Patiente âgée de 58 ans atteinte

d´un cancer du sein (pT1c pN3a cM0; OP 03/2010)

Page 48: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

0

1

2

3

4

5

6

7

Focal Area Background

SU

V M

ax 06.03.12

04.09.12

18.02.13

Suivi longitudinal

Bisphosphonates

Page 49: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Patient âgé de 52 ans atteint d´une néoplasie

neuroendocrine du pancréas (G2 cM1 cN1)

Scintigraphie à la Tc-99m-octéotide 03/2012

Plusieurs foyers hépatiques, douleurs abdominales

Traitement avec 5.1 GBq Y-90-DOTATATE 09/2012

Page 50: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Examen de suivi 4 mois plus tard

Patient âgé de 52 ans atteint d´une néoplasie

neuroendocrine du pancréas (G2 cM1 cN1)

Page 51: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Régression: Continuation de la radiothérapie vectorisée

0

2

4

6

8

10

12

14

16

18

20

Focal Area Background

SU

V M

ax

06.03.2012

21.01.2013

Patient âgé de 52 ans atteint d´une néoplasie

neuroendocrine du pancréas (G2 cM1 cN1)

Page 52: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Quantification TEMP/TDM

Introduction

Quantification

Correction des artefacts

Calibration

Validation

Potentiel clinique

Résumé

Page 53: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Résumé I

La TEMP/TDM quantitative est devenue

accessible au clinicien.

Les erreurs de la mesure de la C* sont < 10 %

dans des volumes > 2 x FWHM.

Page 54: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

Le diagnostic des maladies

affectant l´accumulation des

radiotraceurs d´une manière

plutôt diffuse:

Ostéoporose, neurodégenération,

maladie coronaire, métastases

disséminées

SUV avg 6.2

Résumé II: Potentiel clinique

Page 55: Quantification TEMP/TDM - SFMN SAINT... · 2017-09-12 · SUV avg 6.2 . Quantification TEMP/TDM Introduction Quantification Correction des artefacts Calibration Validation Potentiel

L´évaluation de la réponse des

tumeurs à la thérapie

Dosimétrie de la radiothérapie

moléculaire

(Lu-177, I-123?, I-131?)

SUV avg 6.2

Résumé III: Potentiel clinique